Products: RESPIRATORY CARE: CPAP-BILEVEL S&ST

  • MORITZ S & ST

    The BiLevel Systems for Respiratory Therapy and Non-Invasive Home Ventilation

    Technical details

    Product class IIa according to MPG.

    Mains Connection 100 V AC to 240 V AC, 1 A AC to 0.75 A AC, 50 Hz / 60 Hz.

    Automatic voltage and frequency recognition.

    Electrical Safety Precautions 2 fuses 230 V, 1AT.

    Protection class II, type B.

    Mean Power Consumption Stand-by operation: 13 VA (W).

    Air humidifier off: 29 VA (W).

    Air humidifier on: 40 VA (W).

    Measured under normal breathing at 9 hPa, mean heating power of the air humidifier and 230 V.

    Pressure Range IPAP: 4-18 hPa (mbars) adjustable in 0.1 mbar increments.

    EPAP: 4-18 hPa (mbars, adjustable in 0.1 mbar increments.

    Pressure Fluctuation 4 to 10 hPa: ± 0.25 hPa 10 to 18 hPa: ± 0.5 hPa.

    Measured according to EN ISO 17510-1 with IPAP = EPAP = 6 / 10 / 12 / 18 hPa and VT = 500 ml with 10 / 15 / 20 sinusoidal waves per minute without air humidifier.

    Level of Acoustic Pressure Maximum 30 dB(A).

    Measured according to ISO 17510-1 at 10 hPa in the patient/s vicinity over an angle of 180° in relation to the device front at 10 positions at a distance of 1 m and the same distance from one another at a height of 60 cm.

    Air Filter Filter class G4 (EU4) according to DIN EN 779.

    Flow Delivery Rate 185 l/min on average.

    Dimensions With air humidifier (L x W x H) : 330 mm x 183 mm x 190 mm.

    Weight 4.2 kg without air humidifier 4.5 kg with air humidifier.

    Capacity of the Air Humidifier At least 90 % RH in the breathing mask at 23 °C ambient temperature and 60 % surrounding humidity, 10 hPa, 26 l/min and maximum humidifier power.

    Operating Temperature 10 °C to 35 °C (50 °F to 95 °F).

    Operating Humidity 10 % RH to 90 % RH.

    Operating Air Pressure 800 hPa to 1060 hPa.

    Transport and Storage Temperature -25 °C to + 70 °C (-13 °F to + 158 °F).

    Transport and Storage Humidity 10 % RH to 95 % RH.

    Transport and Storage Air Pressure 700 hPa to 1060 hPa.

    Functions 2-pole mains switch for turning off stand-by operation.

    Liquid crystal display for checking the settings.

    Sleep ramp adjustable from 0 to 45 minutes in 5-minutes increments.

    Automatic start/stop control on/off.

    Keyboard for menu selection and setting parameters.

    Deactivatable acoustic leakage alarm.

    Remote control connection for the physician for setting and displaying the parameters (Twister remote®).

    Display of the respiratory flow curve on the remote control device (Twister remote®).

    Trigger sensitivity adjustable in 5 increments (Twister remote®).

    Pressure rise and pressure drop adjustable in 3 increments (Twister remote®).

    Maximum inspiration time: 0.5 to 12 s, adjustable in 0.1-second increments (Twister remote®).

    Selectable display language G, GB, F, NL, E, I (Twister remote®)

    Additional Function of Moritz®ST Minimum frequency of 4 to 30 breaths per minute, adjustable in increments of 1 breath (Twister remote®).

    Depiction of the inspiration portion of a breath from 20 % to 80 % in increments of 1 % (Twister remote®).

    Change between ST and S mode (Twister remote®)