Products: RESPIRATORY CARE: CPAP-BILEVEL S&ST
- PRESSURE VOLIME VENTILATORS
- CPAP-BILEVEL S&ST
- OXYGEN CONCENTRATORS
- NEBULIZERS
- ASPIRATORS ACTIONS PUMPS
- PEAK FLOW METERS
- SPACERS
- RESPIRATORY TRAINERS
- MASKS-DISPOSABLES
MINI MAX-MAP

The respiratory therapy device with the with - without – with-out concept .
Technical detailsMains Connection 100 V AC to 240 V AC 2 A AC to 1 A AC 50 Hz / 60 Hz
Automatic voltage and frequency recognition.
Electrical Safety Features 1 internal fuse, 230 V, 3.15 AT, Protection class II, type BF
Mean Power Consumption
Stand-by operation: 7 VA (W)
Humidifier off: 16 VA (W).
Humidifier on: 37 VA (W).
Measured under normal breathing at 9 hPa, average humidifier heat output, and 230 V.
Pressure Range 4 to 18 hPa, adjustable in increments of 0.1 hPa.
Pressure Stability 4 to 10 hPa ± 0.5 hPa 10 to 18 hPa ± 5%.
Pressure Fluctuation 4 to 10 hPa ± 0.25 hPa 10 to 18 hPa ± 0.5 hPa.
Measured in accordance with EN ISO 17510-1 with P = 6 / 10 / 12 / 18 hPa at VT = 500 ml with 10 / 15 / 20 sinusoidal cycles per minute.
Noise Level Maximum 30 dB(A).
Measured in accordance with EN ISO 17510-1 at 10 hPA in the patient environment at an angle of 180° to the front of device from 10 positions at a distance of 1 m, at equal distances from one another and at a height of 60 cm..
Air Filter Filter class G4 (EU4) in accordance with DIN EN 779.
Flow Delivery Rate Typically 170 l/min.
Weight 2.3 kg without humidifier 2.9 kg with humidifier.
Humidifier Performance 93% RH in the mask at an ambient temperature of 23 °C and 60 % ambient relative humidity, 10 hPa, 26 l/min and maximum humidifier performance.
Water Consumption Approx. 25 ml H2O/h at an ambient temperature of 23 °C and 60 % ambient relative humidity, 10 hPa, 22 l/min and maximum humidifier performance.
Operating Temperature 10 °C to 35 °C.
Operating Humidity 10 % RH to 90 % RH.
Operating Air Pressure 700 hPa to 1060 hPa.
Transport and Storage Temperature - 25 °C to +70 °C.
Transport and Storage Humidity 10 % RH to 90 % RH.
Transport and Storage Air Pressure 700 hPa to 1060 hPa.
Functions Liquid crystal display for the following unit functions: therapeutic pressure, operating hours, sleep ramp function, date and time.
Adjustable sleep ramp function from 0 (off) to 45 minutes in five-minute increments.
Automatic start/stop control yes/no.
Mask test to check mask fit.
Mains switch to turn stand-by function off.
Humidifier can be connected and disconnected.
Standards DIN EN 60601-1.
General specification for the safety of medical electric devices.
DIN EN 60601-1-2.
Supplement: electromagnetic compatibility.
EN ISO 17510-1.
Sleep apnea therapy device.
- Back