Products: RESPIRATORY CARE: CPAP-BILEVEL S&ST

  • MAGELLAN

    The intelligent discoverer on the way to restful sleep.

    Technical details

    Product class IIa according to MPG (Medizinproduktegesetz = German law of medical device).

    Mains Connection 100 V AC to 240 V AC, 2.2 A AC to 1 A AC, 50 Hz / 60 Hz, Automatic voltage and frequency recognition.

    Electric Safety Precautions 1 internal fuse 230 V, 3.15 AT, protection class II, type B.

    Mean Power Consumption Stand-by operation: 11 VA (W).

    Air humidifier off: 27 VA (W).

    Air humidifier on: 37 VA (W).

    Measured under normal breathing at 9 hPa, average heating capacity of the air humidifier and 230 V.

    Pressure Range 4 to 18 hPa (mbar).

    adjustable in increments of 0.1 hPa.

    Pressure Stability 4 to 18 hPa (mbar) 0.5 hPa 10 to 18 hPa (mbar) 5%.

    Pressure Fluctuation 4 to 10 hPa (mbar) 0.25 hPa 10 to 18 hPa (mbar) 0.5 hPa.

    Measured according to EN ISO 17510-1 with P = 6 / 10 / 12 / 18 hPa at VT = 500 ml with 10 / 15 / 20 sinusoidal waves per minute - variable maximum pressure 6 - 18 hPa (mbar) defines the highest therapy pressure in APAP mode - variable minimum pressure 4- 10 hPa (mbar) defines the lowest pressure in APAP mode - variable APAP pressure 4 – 10 hPa (mbar) defines the initial pressure in APAP mode at turbine start when the sleep ramp is deactivated variable initial pressure 4 – 10 hPa (mbar) defines the initial pressure in APAP mode and in CPAP mode when the sleep ramp is activated.

    Level of Acoustic Pressure Maximum 30 dB(A).

    Measured according to EN ISO 17510-1 at 10 hPa in the patient environment at an angle of 180° in relation to the front side of the device from 10 positions at a distance of 1 m and an equal distance to one another and at a height of 60 cm.

    Air Filter Filter class G4 (EU4).

    According to DIN EN 779.

    Flow Delivery Rate Typically 185 l/min.

    Dimensions With air humidifier (L x W x H): 330 mm x 183 mm x 190 mm.

    Weight Approx. 4 kg.

    Performance of the Air Humidifier At least 90% relative humidity in the mask at 23°C ambient temperature and 60% surrounding humidity, 10 hPa, 26 l/min and maximum humidifier performance.

    Water Consumption Approx. 18 ml H2O/h at 23 °C ambient temperature and 60% surrounding humidity, 10 hPa, 26 l/min, and maximum humidifier capacity.

    Operating Temperature 10 °C to 35 °C.

    Operating Humidity 10 % RH to 90 % RH.

    Operating Air Pressure 800 hPa to 1060 hPa.

    Transport and Storage Temperature -25 °C to +70 °C.

    Transport and Storage Humidity 10 % RH to 90 % RH.

    Transport and Storage Air Pressure 700 hPa to 1060 hPa.

    Functions 2-pole mains switch for turning off stand-by operation .

    Start-/stop button (front / rear) .

    Automatic start/stop control on/off.

    Adjustable sleep ramp from off to 30 minutes in 5-minute increments.

    Mask test.

    Selectable display language (G / GB / F / NL / E / I / TK).

    Keyboard for menu selection and setting parameters.

    Liquid crystal display for checking all settings (Sleep Ramp, automatic start/stop, language, time, date, operating hours, breathing hours, next due for service, mode of operation, maximum pressure, minimum pressure, initial pressure, APAP pressure, therapy pressure, software, serial number).

    Remote control connection for the physician for setting and displaying the parameters (Twister remote®).

    Display of the respiratory flow curve on the remote control device (Twister remote®).

    Connection for Signal Connection Box minni Max nCPAP® to transmit analogue signals of pressure and flow to sleep laboratory systems.

    Data memory for external analysis of the last night and of all events of the last 160 days with the LOGSoft® software.

    Air humidifier functions .

    Hotplate on/off switch.

    Continuous temperature regulator for setting the hotplate temperature

    Scope of delivery Air hose 1.8 m, power cord 2-pole, emergency document, 2 filters, operating instructions, hygiene box.

    LOGSoft® Analysis software LOGSoft® presents the respiratory raw data of the last therapy night as well as all respiratory events over the past 1 to 160 days in a graphical and tabular from that is easy to understand.

    Standards DIN EN 60601-1.

    General specification for the safety of medical electric device.

    DIN EN 60601-1-2.

    Supplement: Electromagnetic compatibility.

    EN ISO 17510-1.

    Sleep apnea therapy device.